Radiation Survey and Recordkeeping Requirements

Contamination Monitoring

Contamination control is a primary emphasis of the radiation safety program at Northeastern University. Prudent contamination control involves careful planning, process containment, protective coverings, personal protective equipment, and avoidance of spills. However, contamination monitoring is the most important element of a contamination control program. Routine contamination monitoring at NU is accomplished through a variety of contamination surveys.

Daily Monitoring

Daily monitoring is to be done anytime a Radioactive Material (RAM) experiment is performed. A portable survey meter is acceptable as long as the sensitivity of the instrument allows for efficient detection of the radionuclides used. See chart at the bottom of this Fact Sheet.

Click on probe type for instructions on how to use the meter. 44-88, 44-9, 44-7or 44-3

A daily survey should cover the hands, lab coat, shoes (top and bottom), work area, floor adjacent to the work area, floor of exit doorways, exit door knobs, and potentially contaminated equipment.

[Top of Page]

Weekly Survey

A thorough contamination wipe survey of the lab is required to be performed every week in addition to daily monitoring during experiments. This weekly survey is to be completed even if NO work is done during the week and must be recorded in DPMs on the area survey form.

A minimum of five locations must be surveyed, including at least one wipe on the floor near the doorway. Wipe locations should include areas and items that have a higher potential for contamination. Locations should be varied from week to week. Some suggested areas to wipe for contamination are:

Wipes must be counted by an appropriate counting technique (LSC or gamma counter). To analyze a wipe with a survey meter click on probe type. 44-88, 44-9, 44-7 or 44-3

Weekly surveys (whether analyzed with a LSC, gamma counter or survey meter) must be documented on an area survey form in DPMs.

If a wipe test reveals contamination exceeding 200 DPM/100 cm2, decontaminate the area and test again. Be sure to fill out a Spill Documentation Form.

Note: Analyzing a wipe with a survey meter can only be performed with approved isotopes. See chart at the bottom of this Fact Sheet.

[Top of Page]

Recordkeeping Requirements

All labs are required to have a Radiation Program binder. The following paperwork must be kept in the binder:

Inventory Sheet

Updated Lab diagram

Area Survey Form

[Top of Page]

Signature Requirements

Daily. Each day RAM is used. A laboratory worker must conduct a survey, document the results and initial the survey records.

Weekly. PI or AI must sign the survey form once a week. “No work” must be written for all days no RAM work was done.

The signature documents that the PI/AL reviewed both the inventory and survey records and resolved any issues with the laboratory staff.

Monthly. PI must sign the survey form at least once a month. The signature documents that the PI reviewed both the inventory and survey records and resolved any issues with the laboratory staff.

[Top of Page]

Inactive Status

If a RAM experiment is not planned in the near future the PI can place the lab in “inactive” status once EHS has the isotopes secured under their supervision. If there is no inventory in the laboratory, there are no documentation requirements. Contact EH&S x2769 or ehs@neu.edu.

[Top of Page]

Survey Instrument Requirements

Isotope

Daily-When RAM is used

Weekly-Done regardless of work

Probe Requirements

Recorded in mR/hr

Counter/Meter Requirements

Recorded in DPMs

P-32, P-33, S-35, In-111, Tc-99m, Cr-51

Geiger-Mueller (G-M) detector

44-9, 44-88

Liquid Scintillation Counter (LSC) or survey meter

Procedures for Analyzing a Wipe Using a Meter

I-125

NaI Solid Scintillation detector

44-3

Gamma counter or survey meter.

C-14

Mueller (G-M) detector

44-9,44-88

LSC

H-3

LSC

Record results in DPMs

Cannot detect H-3 with a survey meter.

LSC

When using a portable survey meter, a count rate of more than two times the background typically indicates the presence of contamination.